Note:
Error messages, calibration see Chapter 16.5.
7.12.8.1 Calculating SPAN value
SPAN indicates the equivalent input voltage in mV/V related to the maximum load (Max)
of the scale.
It is calculated as follows:
* load cell sensitivity C
n
= rated output C
n
(see technical data of the load cell)
The input voltage in mV/V equivalent to the dead load can be calculated by using the
dead load rather than the maximum load using the formula specied earlier.
1. Calculating SPAN value for Max and, if necessary, for the dead see Chapter 7.12.8.1.
2. Press the [by mV/V] softkey.
3. Entering the SPAN value for Max and, if necessary, for the subsequent correction of
dead load (see Chapter 7.12.10).
4. Conrm the entries.
Weight value, weight unit and measuring signal in mV/V corresponding to this
value are displayed in the line [Calibrated at].
The verication is displayed by "Setting Span mV/V".
7 Commissioning Transmitter in eld housing PR 5230
EN-133 Minebea Intec