Procedure:
1. Press the [Linear.] softkey.
The menu linearization is shown.
2. Press the [Add] softkey to set a linearization point.
The input window opens.
3. Enter the desired value using the keyboard.
4. Press the [Ok] softkey.
5. Repeat these steps to set up to three linearization points in succession.
7 Commissioning Transmitter in eld housing PR 5230