[Address]
Enter device address, e.g. for printout.
Input: AZ
[PIN]
The access code can be used to protect the system setup from unauthorized operation.
Input: number with up to 6 digits
As long as you are in this menu, the value can be overwritten as required.
If the [PIN] is set to "0", no access code prompt is displayed.
Note:
SUPER PIN
If the PIN-Code is lost, the setup can be unlocked with Super-PIN "212223."
[Sequence number]
The sequence number (counter for print jobs) is incremented automatically (max.
999999999) and can be set to a start value here as appropriate.
The sequence number can also be shown on the printout (selectable).
Note:
This menu item is only available if under -[Operating parameter]- [Application]
"Standard" has been selected.
[SetTareKey]
Selection: disabled, tare & reset tare, tare & tare again
The tare function (operation via VNC/Internet Browser) can be toggled:
- [tare & reset tare] means that the device will be tared if it has not been tared
previously and the tare will be reset if the device has already been tared.
- [tare & tare again] means that each time a tare command is given, the instant value in
the tare memory is applied and the net display switches to 0.
The tare key has no function when set to [disabled].
[SetZeroKey]
Selection: disabled, only when not tared, reset tare on zeroset
The function of the zero-setting key (operation via VNC/Internet Browser) can be
restricted with [only when not tared] to the gross mode or automatically toggle with
[reset tare on zeroset] to the gross mode.
If the zero-setting key with these settings has no eect, the congured zero-setting
range (around the zero-point set with the dead load) is already utilized due to a previous
zero-setting operation and/or automatic zero setting.
The setting to zero is deactivated by [disabled].
Press to return to the Setup menu.
The following prompt window appears if parameters were changed.
7 Commissioning Transmitter in eld housing PR 5230
EN-187 Minebea Intec