For the conguration the following order must be followed:
1. Dene limits.
2. Dene an action.
3. Determine a condition.
4. Save parameters.
7.15.8.1 Dening limits
Example 1:
The output signal (Limit 1 out) of limit 1 switches OFF above a weight (Wgt) of 900 kg.
The output signal (Limit 2 out) of Limit 2 switches OFF below 290 kg.
The two limit values have a hysteresis of 10 kg.
In the event of a power failure both outputs turn to "o" ("OFF"), thus indicating underll
and overll simultaneously.