The weight or data information requested by the master is transmitted to the master by
PR 5230.
The PR 5230 has write access, the master has read access.
Byte Name Description
0 Read_Value (MSB) e.g., gross value
1 Read_Value (MSB) e.g., gross value
2 Read_Value (MSB) e.g., gross value
3 Read_Value (LSB) e.g., gross value
4 Read_Value_Selected e.g.gross
5 Write
Active
Power
Fail
Out 3 Out 2 Out 1 Limit 3 Limit 2 Limit 1 Status
6 Cmd
Busy
Cmd
Error
Inp. 3 Inp. 2 Inp. 1 Tare
Active
Cal Ac-
tive
Test
Active
Command status
7 Dim-
med
Stand-
still
Inside
ZSR
Center
Zero
Below
Zero
Over-
load
Above
Max
ADC
error
Device status
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Variable Function
Read_Value The weight value is transmitted as a binary 32-bit value with plus or minus sign.
Type of data: DINT
Read_Value_Selected Acknowledgment of the transmitted value.
Write_Active The function selected with Write_Value_Select is executed once.
This bit is deleted if Write_Value_Select is set to 0.
Power_Fail Is set when switching on the device.
Is reset by ResPower with transition from 0→1.
Cmd_Busy The device is busy executing a command (e.g., the device has received a tare com-
mand and is waiting for standstill).
Cmd_Error The device has interrupted the execution of a command (e.g. standstill could not
be reached within the dened standstill time).
The error number can be read at Last error. It is set only if an action is executed.
Tare_Active The scale has been tared.
Cal_Active The device is/was congured.
When this bit is 1, the scale parameters (EXPO/UNIT/STEP) must be read again.
Set after "Power on" and reset after reading the FSD.
Test_Active The device executes the ADC test. The read weight value is not the gross value, but
the test value.
Out Below zero or above Max (FSD).
Standstill The scale is stable.
InsideZSR The weight value is within the zero setting range.
CenterZero The weight value is within center zero (0 ±0.25 d).
Transmitter in eld housing PR 5230 12 Fieldbus interface