EasyManua.ls Logo

Minebea Intec PR 5230 - 4.6.9 ProfiNet I;O interface

Minebea Intec PR 5230
303 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Legend
No. Description
1 Load cell junction board connector
2 Load cell/junction box/scale connector
3 CAL switch
Note:
The weighing electronics board (W1) is also available as a spare part; see Chapter 18.1.1.
Technical data
Description Data
Connection 7-pin plug (male), screen connection on housing and to GN-
DA
Number of channels
1× 6wire or 4wire
Load cell type
Strain gauge, 6 or 4wire connection possible
Load cell supply U = 12 V DC, symmetrical to zero (U = ±6 V DC, I
max
= 160 mA)
Capacity (number of load
cells)
≥75 , corresponding to 8 load cells with 650 or 4 load
cells with 350
Sense input ±6 V DC, with monitoring
Sense voltage monito-
ring
Sense voltage below 8 V, switchable to <8 V
Measurement input
036 mV DC, symmetrical to 0
Dead load suppression max. 36 mV DC (dead load + range)
Accuracy* 0.8 µV/e corr. to 4.8 mV for 6,000 e
0.8 µV/e corr. to 8.0 mV for 10,000 e
Class III, according to OIML R76/EN45501
Min. measuring signal
(OIML)*
6,000 e: ≥0.40 mV/V
10,000 e: ≥0.66 mV/V
Max. resolution 7.5 million internal counts at 3 mV/V
Linearity* <0.003%
Zero standstill error
(TK
0
)*
<0.05 µV/K RTI; 0.004%/10 K at 1 mV/V
SPAN stability error
(Tk
span
)*
< ±2.5 ppm/K
Cable type
6wire with screen for entire cable and screen for measure-
ment cables, e.g. PR 6135/..
Cable length max. 300 m
* with a measurement time of 160 ms .