Conguration PR 5230 PR 5110 conguration
- [Serial ports parameter]- [Remote display]-
[Builtin RS232]- [Param]- [Mode]- [single trans-
mitter]
[ ]- [oP 10]- [LInE]- [rS232]
[ ]- [oP 12]- [tokEn]- [oFF]
[ ]- [oP 13]- [SEndModE]- [SEnd]
[ ]- [oP 14]- [WEIght]- [FolloW]
[ ]- [oP 15]- [WPkEy]- [SElEct]
4.4.5.2 Connecting a YDP14IS ticket printer
The YDP14IS-OCEUV ticket printer can be connected via the internal RS-232 interface.
Conguration PR 5230 Printer conguration
- [Serial ports parameter]- [Printer]- [Builtin
RS232]- [Param]:
- [Protocol] to "RTS/CTS"
- [Baudrate] to "9600"
- [Bits] to "8"
- [Parity] to "none"
- [Stopbits] to "1"
- [Output mode] to "Raw"
The printer must be set to "Line Mode" (the fac-
tory setting is Page Mode). Press the "FEED" key
to switch from one to the other.
The procedure can be found in the printer opera-
ting instructions.
4 Device installation Transmitter in eld housing PR 5230
EN-41 Minebea Intec