Chapter 13 ANALOG OUTPUT
13.2 Analog Output Range
172 HORIBA
(*1) Cell constant 0.00 to 19.99 at 1000m-1
(*2) Cell constant 0.000 to 1.999 at 100m-1
(*3) Cell constant 0.0 to 199.9 10cm-1
(*4) Cell constant 0.00 to 19.99 at 1cm-1
(*5) Cell constant 50.0 to 199.9 at 10cm-1
(*6) Cell constant 5.00 to 19.99 at 1cm-1
(*7) Cell constant 5.00 to 19.99 at 0.1cm-1
(*8) Cell constant 0.500 to 1.999 at 1cm-1
Resistivity 0.200 to 1.999
MΩ•m
1500 to 2000 mV Measurement value × 2.5 ×
10
-4
+ 1500
In Ω•m
20.0 to 199.9
kΩ•m
1000 to 1500 mV Measurement value × 2.5 ×
10
-3
+ 1000
2.00 to 19.99
kΩ•m
500 to 1000 mV Measurement value × 2.5 ×
10
-2
+ 500
0.200 to 1.999
kΩ•m
0 to 500 mV Measurement value × 2.5 ×
10
-1
20.0 to 199.9
(*5)
Ω•m
-500 to 0 mV Measurement value × 2.5 ×
2.5
- 500
2.00 to 19.99
(*6)
Ω•m
-1000 to -500 mV Measurement value × 2.5 ×
25
- 1000
0.50 to 1.999
Ω•m
-1500 to -1000 mV Measurement value × 2.5 ×
250
- 1500
Resistivity 20.0 to 199.9
MΩ•cm
1500 to 2000 mV Measurement value × 2.5 ×
10
-6
+ 1500
In Ω•cm
2.00 to 19.99
MΩ•cm
1000 to 1500 mV Measurement value × 2.5 ×
10
-5
+ 1000
0.200 to 1.999
MΩ•cm
500 to 1000 mV Measurement value × 2.5 ×
10
-4
+ 500
20.0 to 199.9
kΩ•cm
0 to 500 mV Measurement value × 2.5 ×
10
-3
2.00 to 19.99
(*7)
kΩ•cm
-500 to 0 mV Measurement value × 2.5 ×
10
-2
- 500
0.200 to 1.999
(*8)
kΩ•cm
-1000 to -500 mV Measurement value × 2.5 ×
10
-1
- 1000
50.0 to 199.9
Ω•cm
-1500 to -1000 mV
Measurement value × 2.5
-
1500
Tempera-
ture
0 to 100.0 ℃
0 to 1000mV
Measured
object
Measured value Unit Output Conversion method