3 Patient Alarms and INOPs
69
Check Flex Texts
INOP tone
Check the names of the monitor menus, for example the labels for screens, profiles, event or trend
group names, before you resume monitoring. If they are unexpected, there may be a problem with the
monitor software. Contact your service personnel.
Check Keyboard
INOP tone
Perform a visual and functional check of the keyboard. Contact your service personnel.
Check Main Board 2
INOP tone
There is a problem with the second main board in the monitor. Contact your service personnel.
Check Monitor Func
INOP tone
Potential problem with alarm lamps, display or interfaces detected. Contact your service personnel.
Check Monitor Temp
INOP tone
The temperature inside the monitor is too high. Check that the monitor ventilation is not obstructed.
If the situation continues, contact your service personnel.
Check Mouse Device
INOP tone
Perform a visual and functional check of the mouse input device. Contact your service personnel.
Check MSL Voltage
INOP tone
There is a problem with the voltage of the Measurement Link (MSL). Contact your service personnel.
Check Network Conf
INOP tone
The monitor is receiving network topology information from more than one source, e.g. the Database
Server and an Application Server. Contact your service personnel.
Check Screen Res.
INOP tone
The Screen you have selected uses a resolution which is not supported by the display. The monitor will
show a generic Screen instead until you select a different Screen.
Contact your service personnel if you want the Screen deleted from the Profile(s) to avoid this in
future.
Check Settings
INOP tone
If this INOP appears, check the monitor and patient settings before you resume monitoring. If the
settings are unexpected, there may be a problem with the monitor software. Contact your service
personnel.
Check SpeedPoint
INOP tone
Perform a visual and functional check of the SpeedPoint input device. Contact your service personnel.
Check Touch Input
INOP tone
Perform a visual and functional check of the touch input device. Contact your service personnel.
Check Waves
INOP tone
The options purchased with this monitor may not support the number of waves required to show the
selected Screen, so some waves or high resolution trends are missing from the Screen. Select a
different Screen with fewer waves.
Contact your service personnel if you want the Screen deleted from the Profile(s) to avoid this in
future.
CHK ECG Sync Cable
INOP tone
The ECG Sync is detecting an invalid signal, or the ECG Sync cable is disconnected.
Chk IndepDsp Cable The monitor cannot communicate with the D80 Intelligent Display. Check the MSL coupling cable.
The end with the gray connector must be connected to the Intelligent Display.
Chk MSL Connection
INOP tone
Check that the MSL cable is properly connected. If this is the case, try using another MSL cable, to
check if your cable is defective. If this does not help, the device connected via the MSL cable may be
defective, contact your service personnel.
ECG EQUIP MALF T
Numeric is displayed with a -?-
INOP tone
Contact your service personnel.
The ECG in the Telemetry device is faulty.
Internal.Comm.Malf
INOP tone
There is a problem with I2C Bus communication in the monitor. Contact your service personnel.
MCC Reversed
INOP tone
The MSL coupling cable is reversed. Connect the end with the grey connector to the Intelligent
Display.
MCC Unsupported
INOP tone
An MSL coupling cable has been connected to a device which does not support MSL coupling.
INOP Message, Indication What to do