Troubleshooting ST80i Issues
ST80i Stress Test System Instructions for Use A-3
System Navigation Issues
No response to
keyboard commands.
USB cable connection. Check USB cable connection.
Disconnect and connect the keyboard to
check whether keyboard can be
identified by the system
Cursor does not move
USB cable connection. Check USB cable connection.
Disconnect and connect the mouse to
check whether mouse can be identified
by the system.
Treadmill Issues
Treadmill does not
respond to the On
command from ST80i
Equipment was powered up
in the wrong sequence.
The Treadmill power switch
is in the Off position, or the
treadmill cable is not properly
attached.
The Emergency Stop button
is engaged.
Turn the treadmill off by using the Stop
Treadmill
button. Wait one minute and
turn the power On. Proceed with the test.
Ensure the correct equipment is specified
in the Configuration settings.
Check the treadmill-to-ST80i cable
connection. Turn the Treadmill main
power switch to the On position. The
switch is located at the base of the
treadmill hood, on the left side.
Ensure the correct COM port is
configured.
Reset the Emergency Stop button by
turning it clockwise one quarter turn.
System Error Messages
Error message appears:
No exams are currently
selected
An attempt to access a final report
is made, but no patient is selected
from the list.
Click on the patient name to select it, then
access the file.
Table A-1 Troubleshooting ST80i issues
Problem Cause Solution