Troubleshooting and Contacting the Response Center
A-4 ST80i Stress Test System Instructions for Use
System Error Messages (continued)
Error message appears:
Not all of the selected
exams are reported
An attempt was made to print
or export a final report from
the Archive Manager but the
report has not been saved as
Reported.
A test is labeled Reported only after it is
saved.
Click
Save Report, then export the final
report.
Error message appears:
Not reported exam
An attempt to print a final
report from the Report screen
is made but the report has not
been saved as
Reported.
A test is labeled Reported only after it is
saved.
Click
Save Report, then export the final
report.
The Lead Map display
a red “X” on one or
more connections
One or more leads have failed. Connect the electrodes. Once connected, the
warning message disappears.
An error occurred
during export of
selected exams
One or more leads have failed.
Check Export directory.
Check remote location.
Try pinging remote location.
Error message appears:
No stage selected
In the protocol edit function, an
attempt is made to edit a stage,
but no stage has been selected.
Double-click on the stage to edit. The Edit
window appears, allowing you to change the
stage parameters.
Table A-1 Troubleshooting ST80i issues
Problem Cause Solution