Troubleshooting and Contacting the Response Center
A-6 ST80i Stress Test System Instructions for Use
Printer Issues (continued)
No printing LAN settings may be
incorrect.
Network cable may be loose.
Power cord may be
disconnected.
Printer might not have paper
loaded.
Printer error.
Check to see that the Local Area
Network Connection Properties settings
on the PC are defined correctly
Ensure that a crossover network cable is
used and check connections.
Make sure both printer LAN LEDs
(Amber and green) are lit
Ensure the printer port is enabled.
Check AC power cord connections and
confirm that the Power On indicator is
illuminated.
Check to see that paper has been loaded
and that it has been loaded correctly.
Paper sensor mark is at the bottom right.
No motor drive
Insufficient paper tension
Faulty writer assembly
Faulty circuit board
Check and replace as needed.
Defective paper Thermal paper that is old or that
was improperly stored can cause
light or uneven printing.
Exposure to heat or chemical
vapors can damage the paper.
Test printer using a fresh pack of properly
stored paper.
Blood pressure printout
and report discrepancy
Use of NIBP field to enter new
blood pressure values.
You must enter blood pressure values using
the
BP mmhg field in the Side Panel.
For details on entering blood pressure values,
see “NIBP & SpO2” on page 3-29.
Table A-1 Troubleshooting ST80i issues
Problem Cause Solution

Table of Contents