Troubleshooting ST80i Issues
ST80i Stress Test System Instructions for Use A-7
Data-Entry Issues
Cannot manually enter
patient data
It is possible that the Export HL7
check box has been inadvertently
selected on the
Miscellaneous tab
in the Configuration settings.
For the affected profile(s), do the following:
1 Click Settings, then click the
Miscellaneous tab.
2 Ensure the Export: HL7 check box is clear.
3 Click Save Profile and Exit.
Wireless Patient Module Issues
Packet errors/ signal
drop
Signal on the screen shows a red
dashed line.
Configure the system to automatically hop
the RF channel, or perform an RF Scan to
locate the source of the interference and to
change to the optimal channel for the
wireless patient module. For details about the
wireless settings, see the ST80i Wireless
Patient Module Instructions for Use.
Wireless Patient Module Issues (continued)
No waveform signal on
the screen
With the patient (or simulator)
connected, no waveforms are
displayed; only the red dashed
line and lead-fail indicator appear.
NOTE If neither the signal nor the
red dashed line is present, click the
“X” in the upper right corner of the
screen, and then click OK in the
“Abort Test” dialog box.
Check the ST80i System settings, as well as
the wireless patient module to ensure that the
two units are set to the same channel.
For details about setting the receiving
channel settings, see the ST80i Installation
and Configuration Guide.
For details about the wireless settings, see the
ST80i Wireless Patient Module Instructions
for Use for instructions.
Table A-1 Troubleshooting ST80i issues
Problem Cause Solution