Replacement Procedures
Front Case Replacement
Section Menu Section Contents Procedures Back Index
209
8
Disassembling the Front Case
Refer to Figure 8.20 on p. 208.
To disassemble the front case:
1. Remove the printer assembly as described in Removing the Printer Assembly (A12) (p. 360).
2. Separate the front and rear case halves as described in Disassembling the Case (p. 181).
3. Perform steps 1 through 3 to disconnect cables as described in Removing the Interface PCB (A05) (p. 193). Note that J36 and
J37 do not require removal.
4. Remove the four LCD bracket mounting screws (202253-764) (see Figure 8.18 on p. 204). Discard the screws.
5. Remove the printer control keypad cable from the back of the keypad as described in Removing the Printer Control Keypad/
Interface PCB Cable (W12) (p. 216).
6. Remove the main keypad cable from the back of the keypad as described in Removing the Main Keypad/Interface PCB Cable
(W13) (p. 217).
7. Remove the speed dial assembly as described in Removing the Speed Dial Assembly (W15) (p. 218).
8. Remove the speaker assembly as described in Removing the Speaker Assembly (W17) (p. 224).
9. Remove the therapy connector cable as described in Removing the Therapy Connector Cable (W11) (p. 214).
10. Remove the printer assembly/interface PCB Cable (includes removal of the W19 cable) as described in Removing the Printer
Assembly/Interface PCB Cable (W16) (p. 221).