LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Interface PCB (A05) Replacement
Section Menu Section Contents Procedures Back Index
193
8
Removing the Interface PCB (A05)
Refer to Interface PCB Replacement Figures (p. 190) and Inside Front Case Diagram (p. 189).
To remove the interface PCB (3206815) or V2 interface PCB (3306367) from the front case:
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Disconnect the connectors (from top left, counterclockwise) as follows (see Figure 8.11 on p. 191 and Figure 8.12 on p. 192):
~ J31 – Spread the connector locking tab and disconnect the main keypad interface PCB cable (W13, 3207388-001).
~ J37 – Press the connector locking tab and disconnect the backlight interface PCB cable (W06, 3206992-000).
~ J36 – Disconnect the LCD interface PCB cable (W18, 3206990-001).
~ J32 – Press the connector locking tab and disconnect the printer control keypad interface PCB cable (W12, 3206989-
000).
~ J34 – Press the connector locking tab and disconnect the speaker interface PCB wire (W17, 3009726-03).
~ J35 – Spread the connector locking tabs and eject the printer interface cable (W16, 3009724-001).
~ J33 – Press the connector locking tab and disconnect the cable for the speed dial (W15, 3011128-002).
3. (If replacing the interface PCB) Disconnect the System/Interface cable as described in System/Interface PCB Cable (W04)
Replacement (p. 212).
4. Remove the screw holding the ground strap. Discard the screw.
5. Remove the eight remaining screws (202253-761) (see Figure 8.10 on p. 190). Discard the screws.