LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Front Case Replacement
Section Menu Section Contents Procedures Back Index
210
8
Assembling the Front Case
Refer to Figure 8.20 on p. 208.
To install the new front case: 15 steps, (Page 1 of 2)
NOTE: Use the Front Case Repair Kit (MIN 3305431-023) (p. 511).
NOTE: Transfer the following parts from the old front case to the new front case:
~ LCD/interface PCB assembly
~ W15 speed dial assembly
~ W17 speaker assembly
~ W11 therapy connector
~ W16 printer/interface PCB cable
~ W12 printer control keypad cable
~ W13 main keypad cable
~ Printer bracket (3006810-01) with W19 cable
1. Install the perimeter seal (804234-03) into the groove on the front case.
NOTE: Thread the perimeter seal approximately 1.5 inches through the seal lock.
Do not stretch the seal, as this will result in crimped corners.
2. Install the speed dial assembly as described in Installing the Speed Dial Assembly (W15)
(p. 220).
3. Install the therapy connector cable as described in Installing the Therapy Connector Cable
(W11) (p. 214).
4. Install the printer assembly/interface PCB Cable (includes the W19 cable) as described in
Installing the Printer Assembly/Interface PCB Cable (W16) (p. 222).