LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Front Case Replacement
Section Menu Section Contents Procedures Back Index
211
8
5. Install the speaker assembly as described in Installing the Speaker Assembly (W17) (p.
225).
6. Install the display lens as described in Installing the Display Lens (p. 207).
7. Install the display shield as described in Installing the Display Lens Shield (p. 203).
8. Install the LCD/Interface PCB assembly by installing the LCD mounting bracket in the front
case and secure with four new screws (202253-764); torque to 6.8 in-lb.
9. Reconnect cables to interface PCB by performing step 3 as described in Installing the
Interface PCB (A05) (p. 194).
10. Install the printer control keypad as described in Installing the Printer Control Keypad
(A09) (p. 199).
11. Install the main keypad as described in Installing the Main Keypad (A10) (p. 201).
12. Reassemble the case as described in Reassembling the Case (p. 184).
13. Clean the front case with isopropyl alcohol, and apply Physio-Control logo (3208003-000)
to the location provided in upper left of front case.
14. Install the printer assembly as described in Installing the Printer Assembly (A12) (p. 361).
15. Installing the label set (3207318-XXX, see Table 9.14 on page 425) is required. Clean the
front case areas with isopropyl alcohol and install label numbers 4, 6, 10 and 11.
To install the new front case: (Continued) 15 steps, (Page 2 of 2)