Energy Storage Capacitor (A15) Replacement
Section Menu Section Contents Procedures Back Index
272
8
Removing the Energy Storage Capacitor (A15)
To remove the Energy Storage Capacitor from the rear case:
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the system/therapy PCB assembly as described in System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232).
3. Remove the transfer relay as described in Removing the Transfer Relay Assembly (A13) (p. 269).
NOTE: It is not necessary to remove the spade terminals as described in the procedure.
4. Disconnect the energy storage capacitor (3008164-002) white lead from the interconnect bracket at pin 1. Needle nose pliers
may be used to assist in disconnection.
5. Disconnect the energy storage capacitor red lead from the interconnect bracket at pin 4. Needle nose pliers may be used to assist
in disconnection.
6. Cut the restraining large cable tie (200536-011) from the interconnect bracket.
7. Lift out the interconnect bracket and storage capacitor from the case.