NIBP (A21)/CO2 (A23) Module Replacement
Section Menu Section Contents Procedures Back Index
288
8
Removing the NIBP (A21)/CO2 (A23) Modules
To remove the NIBP and CO2 modules as a unit from the rear case: 17 steps, (Page 1 of 4)
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the system/therapy PCB assembly as described in System (A01)/
Therapy (A04) PCB Assembly Replacement (p. 232).
3. Remove the transfer relay assembly as described in Removing the Transfer
Relay Assembly (A13) (p. 269).
NOTE: It is not necessary to remove the spade terminals as
described in the procedure.
4. Remove the OEM PCB as described in Removing the OEM PCB (A06) (p.
262).
5. Remove the three mounting bracket screws (3207337-312). Discard the
screws.
6. If present, disconnect the CO2 exhaust tube. Lift the assembly as needed.
bracket mounting
screws (3 places)
CO2 exhaust from
rear case
Figure 8.81—CO2/NIBP mounting bracket