LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
CO2 Inlet Connector Cable (W28) Replacement
Section Menu Section Contents Back Index
351
8
Removing the CO2 Inlet Connector Cable (W28)
To remove the CO2 inlet connector cable (3012140-008) from the parameter bezel (rear case—refer to Inside Rear Case Diagrams (p.
228) and Figure 9.21 on p. 413):
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the system/therapy PCB assembly as described in Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233).
3. Remove the parameter bezel as described in Removing the Parameter Bezel (p. 311).
4. Remove the four screws (202253-732) securing the CO2 connector adapter (3012119-02) to the parameter bezel.
5. Remove the CO2 connector adapter and, from the underside of the bezel, remove the CO2 connector retainer (3012121-002)
and seal (3007997-01). Discard the seal and screws.
6. Disengage the CO2 inlet connector cable from the CO2 connector retainer by spreading the arms of the retainer and sliding out
the CO2 inlet connector body (see Figure 8.118 on p. 350). Discard the connector retainer clip.