EasyManua.ls Logo

RKI Instruments GX-6000 - Page 141

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GX-6000 Operator’s Manual PID Sensor Maintenance 132
pushed down in the center and secured under the gasket aligning tabs.
Failure to install the sensor gasket appropriately may result in
inaccurate gas readings.
Figure 35: Sensor Gasket Seating
21 . Insert the flow chamber back into the instrument.
22 . Tighten the flow chamber’s four screws that were loosened in Step 2.
23 . Start up the GX-6000 by pressing and briefly holding the POWER/
ENTER button.
24 . To verify that the sensor gasket was inserted properly:
a. Plug the inlet with your finger.
b. Verify that the GX-6000 goes into low flow alarm. If the GX-6000
does not go into low flow alarm, turn the instrument off and attempt
to seat the sensor gasket again.
25 . Calibrate the PID sensor as described in “Chapter 4: Calibration Mode”
on page 68.
Gasket Aligning Tabs
Flow Fitting Connections

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals