EasyManua.ls Logo

RKI Instruments GX-6000 - Page 207

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GX-6000 Operator’s Manual IR CO
2
Start Up and Normal Operation 198
2 . Connect the tubing to the GX-6000’s inlet fitting. Be sure the arrow on
the filter is pointing toward the inlet fitting.
Figure 42: CO2 Scrubber Connected to GX-6000
3 . Wait one minute for the fresh air sample to flow through the carbon
dioxide scrubber, then press and hold the AIR button to set the zero
reading.
4 . Remove the scrubber from the inlet fitting.
5 . Reattach the tubing to the open end of the scrubber.
Performing a Demand Zero for Instruments with a
Carbon Dioxide Sensor and a PID Sensor
If your instrument has both an IR CO
2
sensor and a PID sensor, you will need
to use both the carbon dioxide scrubber and the VOC zero filter to perform a
demand zero.
1 . Remove the CO
2
scrubber’s tubing from the side that has an arrow
pointing to it.
2 . Remove both plugs from the tubing stubs on the VOC zero filter.
3 . Connect the VOC zero filter to the GX-6000’s inlet fitting. The filter
does not have a preferred flow direction.
4 . Connect the CO
2
scrubber to the tubing stub on the other side of the
VOC zero filter. Be sure the arrow on the CO
2
scrubber is pointing
toward the VOC zero filter. See the figure below for the scrubber setup.
Figure 43: CO2 Scrubber and VOC Zero Filter Combination
5 . Wait one minute for the fresh air sample to flow through the scrubber
setup.
6 . Press and hold the AIR button to set the zero reading.
To GX-6000
CO2 Scrubber
VOC Zero FilterCO2 Scrubber
To GX-6000

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals