209 • Leak Testing GX-6000 Operator’s Manual
3 . Press and hold the SHIFT â–¼ (PANIC) button, then press and hold the
â–²AIR button and hold both until you hear a beep. The Station ID Select
Screen will display and the current station ID will be in the middle of the
screen.
4. To exit Snap Log Mode and return to Leak Check Mode, press and
release the DISP/LOCK button.
5. Use the â–²AIR and SHIFT â–¼ (PANIC) buttons to scroll to the desired
station ID.
6. Press and release the POWER/ENTER button to continue to the Base
Reading Screen.
NOTE: If CO DISPLAY is set to ON, the CO reading is saved during a
snap log but will not be displayed in the Base Reading Screen, Peak
Reading Screen, or while viewing snap log data at the instrument.
However, the saved CO reading(s) can be viewed using the
GX-6000 Datalogging Program.
7. To return to the Station ID Select Screen without taking a base reading,
press and release the RESET button. To exit Snap Log Mode completely
and return to Leak Check Mode, press and release the DISP/LOCK