GX-6000 Operator’s Manual Leak Testing • 212
3 . If snap logs have been taken, the screen indicates the year, month, day,
and time that the most recent snap log was taken.
The number near the bottom of the screen indicates the snap log ID
number. The first snap log that is taken is given an ID of 001. The next
snap log ID is 002. The ID number increases sequentially with each set
of snap log data.
The last line of the screen indicates the Station ID that was used for the
snap log.
4 . Press and release the DISP/LOCK button to return to Leak Check Mode.
5 . Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll through different
snap log IDs and view what time they were taken. Snap log data that was
taken in Leak Check Mode operation will have two snap log ID
numbers. One number is for the gas reading and the other is for the peak
reading.
6. To view the data in a snap log ID, press and release the POWER/ENTER
button. The gas readings that were taken during the snap log are
displayed.