251 Maintenance GX-6000 Operator’s Manual
2 . Remove the used tube. Take care not to touch the broken ends of the
tube.
3 . Break off the tips of the new tube using the tip-breaking feature in the
tube holder. The reservoir can hold approximately 12 tips and will need
to be periodically emptied.
4 . Insert the new tube in the front half of the tube holder. The arrow on the
tube should be pointing away from the front half.
Figure 48: Inserting New Tube Into Front Half
5 . Line up the locking feature in the front half with its mate in the back half
and insert the front half/tube assembly into the back half making sure the
locking feature engages.
Figure 49: Reassembling Tube Holder
6 . If you are replacing your old tube with a tube from a new box, compare
the calibration code listed on the old box to the calibration code listed on
the new box.
a. If the calibration codes match, there is no need to do anything else.
b. If the calibration codes do not match, you must either perform a
TUBE CAL using a tube from the new box (see “Performing a TUBE
CAL” on page 241) or you must enter the new calibration code into
CAL CODE (see “Entering a CAL CODE” on page 248), depending
on what method of calibration you used with your old box of filter
tubes. See “Calibrating the 10.0 eV/Benzene PID Sensor” on
page 239 for a description of calibration methods.
7 . Tubes do not contain cadmium, mercury, or chrome and may be disposed
as general waste.