27 • Start Up GX-6000 Operator’s Manual
9 . The full scale values and the gas alarm setpoints are displayed by five
screens in sequence: the Full Scale Screen, the Low Alarm Screen, High
Alarm Screen, STEL Alarm Screen, and TWA Alarm Screen. Each
screen remains on the LCD for three seconds.
10 . If ID DISPLAY is set to ON (see “Turning the User/Station ID Function
On or Off” on page 160), the User ID Screen appears for a few seconds,
followed by the Station ID Screen.
If ID DISPLAY is set to OFF, these screens will not appear.
F.S.
CH4 O2
%LEL %
H2S CO
ppm ppm
VOC P
ppm
100 40.0
100.0 500
6000
AL1
CH4 O2
%LEL %
H2S CO
ppm ppm
VOC P
ppm
10 19.5
5.0 25
400.0
AL2
CH4 O2
%LEL %
H2S CO
ppm ppm
VOC P
ppm
50 23.5
30.0 50
1000
STEL
H2S CO
ppm ppm
VOC P
ppm
1.0 25
60.0
TWA
H2S CO
ppm ppm
VOC P
ppm
5.0 200
42.0
4:07
USER ID
KIMBERLY
4:07
STATION ID
TANK 5