Roche Diagnostics
Instructions for Use · Version 13.0 271
cobas b 123 POC system 10 Software functions
Configuring network connection settings
p To manually connect your system to a network using an IP address
1
Utilities > Configuration > System > Connectivity > Network
Figure 10-28 Configuring network connection screen
2
In the IP address panel, press the and buttons on the upper part of the
panel to display Manual.
3
To enter an IP, subnet mask and/or gateway address, press the button on their
respective fields. Enter your addresses using the numeric keyboard box that
appears on the screen. Then, press the OK button.
4
To apply your changes, press the Close button.
s
p To manually connect your system to a network using DHCP and DNS
servers
1
Utilities > Configuration > System > Connectivity > Network
Figure 10-29 Configuring network connection screen