12 Troubleshooting cobas b 123 POC system
System stops
15 Instrument error Analyzer defective Fluidics control board
defective
o Press the Shutdown system button.
o Turn on the instrument again.
o If this error persists, contact your Roche Service
representative.
16 Instrument error Analyzer defective Oximeter module board
defective
o Press the Shutdown system button.
o Turn on the instrument again.
o If this error persists, contact your Roche Service
representative.
17 Instrument error Analyzer defective UIM hardware/power
supply defective
o Press the Shutdown system button.
o Turn on the instrument again.
o If this error persists, contact your Roche Service
representative.
18 Sensor Cartridge
error
Sensor Cartridge
dirty
o Press the Change Sensor Cartridge button. Follow
the instructions on the screen.
o Inspect the rear side of the Sensor Cartridge for dirt.
Remove any dirt with a dry cloth.
o Reinstall the Sensor Cartridge.
o If this error persists, remove the Fluid Pack:
Workplace > Change Fluid Pack
o Inspect the docking parts of the Fluid Pack for dirt.
Remove any dirt with a moist cloth.
o Reinstall the Fluid Pack.
o If this error still persists, contact your Roche Service
representative.
19 Fluid Pack error Fluid Pack defective Fluidics tightness
problem
o Press the Change Fluid Pack button. Follow the
instructions on the screen.
o If this error persists, contact your Roche Service
representative.
20 Fluid Pack error Fluid Pack defective Fluidics system blocked
o Press the Change Fluid Pack button. Follow the
instructions on the screen.
o If this error persists, contact your Roche Service
representative.
21 Instrument error Spare part not
initialized
Unknown C board
o Press the Restart system button.
o If this error persists, contact your Roche Service
representative.
22 Instrument error Spare part not
compatible
EEPROM data
conversion error
o Press the Restart system button.
o If this error persists, contact your Roche Service
representative.
23 Instrument error Analyzer defective C power connection
error
o Press the Shutdown system button.
o Turn on the instrument again.
o If this error persists, contact your Roche Service
representative.
24 Sensor Cartridge
chip
communications
error
Sensor Cartridge
chip error
Read/write chip error
o Press the Change Sensor Cartridge button. Follow
the instructions on the screen.
o Inspect the Sensor Cartridge chip for dirt. Remove
any dirt with a dry cloth.
o Reinstall the Sensor Cartridge.
o If this error persists, contact your Roche Service
representative.
ID Category Cause Type of error Action
Table 12-1 System stops