Remote Control Commands
R&S
®
SMA100B
459User Manual 1178.3834.02 ─ 05
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS2:SWE:FREQ:SENS:SRAN ON
Activates use of a separate frequency range for frequency ver-
sus power measurement for sensor 2.
SENS2:SWE:FREQ:SENS:STAR 2.0GHZ
Sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
SENS2:SWE:FREQ:SENS:STOP 2.9GHZ
Sets a sweep stop at 2.9 GHz irrespective of the current signal
generator frequency settings.
Manual operation: See " Use Separate Frequency- Power Analysis " on page 240
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet <Offset>
Defines the level offset at the sensor input in dB. Activate the offset with the com-
mand :SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet:STATe
on page 459.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
SENS2:SWE:POW:OFFS -5dB
Defines a level offset of -5 dB.
SENS2:SWE:POW:OFFS:STAT ON
Activates that the specified level offset is considered.
Options: Option R&S SMAB-K28
Manual operation: See " Level Offset - Power Analysis " on page 240
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet:STATe <State>
Activates a level offset at the sensor input. Define the appropriate value with the com-
mand :SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet on page 459.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS2:SWE:POW:OFFS:STAT ON
Activates the specified level offset.
Options: Option R&S SMAB-K28
Power Sensor Measurement Subsystems