Remote Control Commands
R&S
®
SMA100B
502User Manual 1178.3834.02 ─ 05
Parameters:
<Mapping> SENS1 | SENSor1 | SENS2 | SENSor2 | SENS3 | SENSor3 |
SENS4 | SENSor4 | UNMapped
Sensor channel.
*RST: UNMapped
Example: See Example "Detecting and assigning a power sensor"
on page 499.
Manual operation: See " Sensor Mapping List " on page 213
:SLISt:SENSor:MAP
<SensorId>, <Mapping>
Assigns a sensor directly to one of the sensor channels, using the sensor name and
serial number.
To find out the the sensor name and ID, you can get it from the label of the R&S NRP,
or using the command :SLISt:SCAN[:STATe]. This command detects all R&S NRP
power sensors connected in the LAN or via 'USBTMC protocol.
Setting parameters:
<SensorId> string
<Mapping> enum
Example: See Example "Detecting and assigning a power sensor"
on page 499.
Usage: Setting only
Manual operation: See " Sensor Mapping List " on page 213
:INITiate<hw>[:POWer]:CONTinuous <Continuous>
Switches the local state of the continuous power measurement by R&S NRP power
sensors on and off. Switching off local state enhances the measurement performance
during remote control.
The remote measurement is triggered with :READ<ch>[:POWer]?). This command
also returns the measurement results. The local state is not affected, measurement
results can be retrieved with local state on or off.
Parameters:
<Continuous> 0 | 1 | OFF | ON
*RST: 0
Example:
INIT1:CONT ON
Switches local state of continuous power measurement on.
Manual operation: See " State " on page 218
:READ<ch>[:POWer]?
Triggers power measurement and displays the results.
SENSe, READ, INITiate and SLISt Subsystems