146 Brakes / Movements
SIREMOBIL Iso-C SPR2-230.840.01.01.02 © Siemens, 2008
08.08 CS PS SP
Page 146 of 148
For internal use only
x Repeat the procedure at the other end of the C-arm.
x After the counterbalance weights have been reattached on both sides, move the C-arm
orbitally from one end to the other several times.
x In the case of SIREMOBIL Iso-C 3D, also perform a scan.
x If the belt tension is still insufficient, repeat the procedure and move the belt to the next
tooth to shorten it further.