Prerequisites 19
Page 19 of 148
For internal use only
NOTE
For evaluation purposes, the first measured value and the
values documented during previous maintenance or safety
checks must be compared to the measured values.
A sudden or unexpected increase of the measured values
may indicate that a fault has occurred in the primary circuit
of the power supply (damaged insulation, water or humidity
damage, defective interference suppressor, etc.) - even if the
limit value of 2.5 mA is not exceeded.