x Select "5. CLOSED POSITION" on the service PC and click "Execute".
x Close the status window by clicking the OK button.
x For the query "Do you want to reset the Unit", click "No".
x Select "8. Init Off & Store Positions" on the service PC and click "Execute".
x Close the status window by clicking the OK button.
x For the query "Do you want to reset the Unit", click "Yes".
x Switch the SIREMOBIL Iso-C off and on.
X
x Check the X-Iris again in survey format and zoom format. The collimator blades must
still be visible on the edge of the image. If necessary, repeat the adjustment.
x Return to the main menu on the service PC with the ESC key.
x Select the Data, Backup... menu.
x Select Parameter in the backup window.
x Click "Backup".
x At "Remarks for Parameter" enter "X-Iris set".
x Click "OK" . The new position values for the collimator will be stored on the disk.
x Return to the main menu with the ESC key.