EasyManua.ls Logo

Siemens SIREMOBIL Iso-C

Siemens SIREMOBIL Iso-C
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Adjustments / Programs 67
Page 67 of 148
For internal use only
x After the adjustment, select "Stop Diamentor Test" on the service PC and click "Exe-
cute".
x Close the status window by clicking the OK button.
x Respond to the query "Do you want to reset the unit" by clicking "No".
x The test is finished.
x Check the accuracy of the dose-area product measurement device. See (Checking the
accuracy of the dose-area product measurement device / p. 70).

Table of Contents

Other manuals for Siemens SIREMOBIL Iso-C

Related product manuals