x In the "Available Functions" window select the "Start Diamentor Test" line and click
"Execute".
x Close the popup window by clicking the OK button.
x Close the prompt "Do you want to reset the unit?" by clicking the "No" button.
x Start the test by pressing the ->0<- key (reset key for the fluoroscopic time).
¹ The tested value is displayed on the SIREMOBIL monitor.
x The displayed value must agree with the programmed test value "50". Tolerance: r 5 %.
x If the displayed value is outside tolerance, replace the dose-area product measure-
ment device.
x After the check on the service PC, select "Stop Diamentor Test" and click "Execute".
x Close the popup window by clicking the OK button.
x Close the prompt "Do you want to reset the unit?" by clicking the "No" button.
x The test is finished.
x Check the accuracy of the dose-area product measurement device. See (Checking the
accuracy of the dose-area product measurement device / p. 70).