5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
The DAC UNIVERSAL automatically runs the cleaning,
lubrication and sterilization cycle.
The display shows at any time which phase the DAC
UNIVERSAL is in, see section "Display information DAC
UNIVERSAL ADVANCED [ → 71]“.
When the selected cycle is completed the display will indicate as
seen on the picture.
The lid moves about halfway out of the chamber. This indicates
that a cycle has been completed and the instruments have been
sterilized.
The cycle number is shown next to the completed cycle symbol.
13. Press the "Down"-Key.
The STANDARD lid will open completely.
CAUTION
Danger of burning!
Never hold your head directly over the DAC UNIVERSAL or touch the
inserted lid, while the DAC UNIVERSAL opens, as steam comes out of
the unit upon completion of a cycle.
WARNING
After completion of the cycle, make sure that all instruments are still
properly attached to the STANDARD lid or the adapters. If instruments
have fallen into the chamber during the cycle or if an instrument is no
longer properly attached after the end of a cycle, cleaning and
sterilization have failed. In this case, reprocessing of the instruments
must be repeated.
CAUTION
Danger of burning!
Even when the STANDARD lid is half or completely open, the
instruments are still hot. Allow the extremely hot instruments to cool
down after the process.
CAUTION
Danger of burning!
The metal part of the lid is very hot. Please touch only the plastic part of
the STANDARD lid at the end of the cycle.