Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J B-19
12. Press the Yes button to calculate the new span number, and the
following screen appears. (If you press the Advance button, the
Micro-Tech displays the difference from the previous setting.)
13. Press the Run button to return to the Material Calibration screen
shown below. (If you press the Factor button, the Micro-Tech uses
a simulated calibration to populate the parameters in the Material
Factor function.)
14. Press the No button and the Run screen reappears.
Here is the Material Calibration menu table.
Parameter Selection Description
MAT’L CALIBRATION
no, yes
This scroll is only available
when Linearization scroll 1
is set to Yes.
3 points linear.
MAT’L CALIBRATION
START, MENU
BATCH
SPAN # CHANGED
New span # xxxxxx
Old span # xxxxxx
READY
ALARM
CALIB
MENU
RUN
FACTOR
BATCH
MAT’L CALIBRATION
Add reference
weight to totals?
READY
ALARM
CALIB
NO
YES