Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J B-27
revolutions to pass
reference
Time required for minimum number of
revolutions (seconds).
range +/-
The AZT sequence starts automatically when
the flow rate is below this percentage of the
max scale capacity.
15 Auto zero tracking
step +/-
0.02% (0.01 – 10 )
16 Auto zero tracking
max dev. +/-
4.0% (0-10);
(0-2)*
(0-5)**
*NTEP W&M mode.
**NTEP W&M mode when print AUTO ZERO
is enabled under the PRINT scroll.
Generates an alarm when the change from
the Zero Reference value to the value of zero
with new factor applied exceeds the max
Warm Time
Test duration
Only available when Nr. of test duration is set
at 2.
This scroll is only available when NTEP
W&M mode is selected.
Generates an alarm when the change from
the Zero Reference value to the new zero
value exceeds the max deviation.
The base value for Autozero max deviation
and Autozero Track max deviation. This
value must be re-set to clear a Autozero or
Autozero Track max deviation alarm.