106
MAINTENANCE
USER MAINTENANCE ITEMS
MANUFACTURER MAINTENANCE ITEMS
BRIGHTNESS ADJUSTMENT OF THE CONTROL PANEL
• The control panel is optimally adjusted when shipped.
• For control panel brightness adjustment, see "INITIAL (INITIAL SETTING)," "SETTING
THE CONTROL PANEL BRIGHTNESS" (page 83).
Item Inspection time Contents
Inspection Before using
• The instrument works properly.
• The objective lens must be free of
stains and/or flaws.
• Confirm whether the foreign object is
attached to the measuring nozzle and
the area around the measuring nozzle.
•Air check
• Confirm that the safety stopper setting
and measuring nozzle do not move to
the patient's side beyond the safety
stopper setting position.
Cleaning When the part is stained
• Objective lens
• External cover, control panel, etc.
Replacement As required
•Fuse
• Printer form
Item Checking time Contents
Cleaning each
component
Within 12 months
• Cleaning outer covers
• Checking the optical system
• Cleaning POWER unit
Operation check Within 12 months
• Checking the main body operation
• Checking switches
Accuracy check Within 12 months
• Confirming the ocular pressure mea-
surement functions (using special tools)
• Confirming the cornea thickness mea-
surement functions (using special tools)
(only in CT-1P)