106
MAINTENANCE
USER MAINTENANCE ITEMS
MANUFACTURER MAINTENANCE ITEMS
BRIGHTNESS ADJUSTMENT OF THE CONTROL PANEL
The control panel is optimally adjusted when shipped.
For control panel brightness adjustment, see "INITIAL (INITIAL SETTING)," "SETTING
THE CONTROL PANEL BRIGHTNESS" (page 83).
Item Inspection time Contents
Inspection Before using
The instrument works properly.
The objective lens must be free of
stains and/or flaws.
Confirm whether the foreign object is
attached to the measuring nozzle and
the area around the measuring nozzle.
•Air check
Confirm that the safety stopper setting
and measuring nozzle do not move to
the patient's side beyond the safety
stopper setting position.
Cleaning When the part is stained
Objective lens
External cover, control panel, etc.
Replacement As required
•Fuse
Printer form
Item Checking time Contents
Cleaning each
component
Within 12 months
Cleaning outer covers
Checking the optical system
Cleaning POWER unit
Operation check Within 12 months
Checking the main body operation
Checking switches
Accuracy check Within 12 months
Confirming the ocular pressure mea-
surement functions (using special tools)
Confirming the cornea thickness mea-
surement functions (using special tools)
(only in CT-1P)