EasyManua.ls Logo

Topcon CT-1P - REFERENCE MATERIAL: OPTIONAL ACCESSORIES; OPTIONAL ACCESSORIES LIST; SHAPE OF PLUG GUIDELINES

Topcon CT-1P
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
120
GENERAL INFORMATION ON USAGE AND MAINTENANCE
ELECTROMAGNETIC COMPATIBILITY
The product conforms to the EMC standard (IEC 60601-1-2 Ed3.0:2007)
a)MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS.
b)Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c)The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYS-
TEM as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d)The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e)The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other
than those specified may result in increased EMISSION or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
Item Article code Cable Shielded Ferrite Core Length (m)
AC power cord for EUT 44804 7017 No No 1.5
USB cable
-
Yes No 2.5
LAN cable
-
No No 1.5
Serial cable
-
Yes No 3.0
DC cable for PC AC adapter
-
No No 1.8
AC power cord for PC AC adapter
-
No No 1.8
Guidance and manufacturer's declaration - electromagnetic emissions
The CT-1P/CT-1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CT-1P/CT-1 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The CT-1P/CT-1 uses RF energy only for its inter-
nal function. Therefore, its RF emissions are very
low and are not likely to cause any interference
with nearby electronic equipment.
RF emissions
CISPR 11
Class B
The CT-1P/CT-1 is suitable for use in all establish-
ments other than domestic and those directly con-
nected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies

Table of Contents