GENERAL INFORMATION ON USAGE AND MAINTENANCE
PATIENT'S ENVIRONMENT
When the patient or inspector may touch the devices (including the connecting devices) or
when the patient or inspector may touch the person that comes into contact with the devices
(including the connecting devices), the patient's environment is shown below.
In the patient's environment, use the device conforming to IEC60601-1. If you are compelled
to use any device not conforming to IEC60601-1, use an insulation transformer or the com-
mon protective earth system.
Note 1: Use the device conforming to IEC60950-1.
CAUTIONS
• Don't connect an additional power strip or an extension cord
to the system.
• Don't connect the device which is not recognized as one
component of the system.
Radius of 1.5m
2.5m
1.5m1.5m
Do not use the power strip
in the patient's environ-
ment. Connect the power
supply of the device to the
commercial power supply.
Devices applicable
to the patient's
environment
• Barcode reader
(Note 1)