SETTING FUNCTIONS ON THE SETUP SCREEN
LIST OF SETUP ITEMS
Setup items are categorized into 6 large indexes.
"Initial" .....................items related to the initial status after power on
"Print" ......................items related to output from the internal printer
"Comm" ...................items related to data output with the external device
"LAN".......................items related to output using the LAN
"Operator ID"...........items related to Operator ID.
"Special" ..................items related to maintenance (for service engineer only)
INITIAL (INITIAL SETTING)
Initial contains settings related to the initial status after power on, clearing all measure-
ment values, etc.
Descriptions Options Details Initial value Page
Buzzer
OFF Buzzer does not sound.
ON 68
ON Buzzer sounds.
Start mode
MANUAL Default measurement mode is MANUAL.
Auto 69
AUTO Default measurement mode is AUTO.
Auto print
ON
After AUTO measurement,
results are printed out automatically.
ON 69
OFF Not printed automatically.
Printer
OFF Internal printer is disabled.
ON 69
ON Internal printer is active.
Show
Tono value
OFF Tono value is not displayed.
ON 70
ON Tono value is displayed.
Show
Pacho value
*
1
OFF Pacho value is not displayed.
ON 70
ON Pacho value is displayed.
Show
Tono average
OFF Tono average value is not displayed.
ON 71
ON Tono average value is displayed.
Show
Adj value
*1
OFF Adjusted value is not displayed.
ON 71
ON Adjusted value is displayed.
Patient No.
reset
OFF Patient No. is not reset upon power on.
OFF 72
ON Patient No. is reset upon power on.
Show patient
ID
OFF Patient ID is not displayed.
ON 72
ON Patient ID is displayed.
Required
patient ID
OFF Patient ID is not displayed.
OFF 73
ON Patient ID is displayed.
Device ID
number
1-99 Set by
ten-key display
Sets the Device ID number. 1 73
Show Device
ID number
OFF Patient ID is not displayed.
OFF 74
ON Patient ID is displayed.
Stand by
mode
RIGHT Waiting at the initial position for right eye measurement.
RIGHT 74LEFT Waiting at the initial position for left eye measurement.
LAST Waiting at the last position of the measured eye.
Auto power
save
OFF Power save function is not used.
10min 75
1min Power save status in 1min after last operation.
5min Power save status in 5min after last operation.
10min Power save status in 10min after last operation.
20min Power save status in 20min after last operation.
30min Power save status in 30min after last operation.
60min Power save status in 60min after last operation.