5 Place the patient's chin on the chinrest and check that his/her forehead is touching to the
forehead rest.
6 Press the button to adjust the chinrest height until the eye height mark of the
chinrest reaches the same height as the patient's eye.
SETTING THE SAFETY STOPPER
1 Select the right/left eye by tapping the button/ button.
CAUTIONS
• Before measuring, set the safety stop to prevent the mea-
suring window glass from hitting the patient's eye.
Set it respectively for the right and left eyes.
• Set the safety stop from the side of the instrument.
Setting operations from other positions, where checking of
the eye position is not easy, may cause injury.
Eye height mark
Up/down button
for chinrest