a<CR><LF>
a
Single-digit WMT700 address that corresponds to the first character of the address
value
<CR><LF>
Terminates the response
5.26.1.2 a! — Acknowledge Active
This command ensures that WMT700 is responding to a data recorder or another SDI-12
device. It asks WMT700 to acknowledge its presence on the SDI-12 bus.
a!
a
Single-digit WMT700 address that corresponds to the first character of the address
value
!
Terminates the command
The response:
a<CR><LF>
a
Single-digit WMT700 address that corresponds to the first character of the address
value
<CR><LF>
Terminates the response
1!1<CR><LF>
5.26.1.3 aAb! — Change Address
This command changes the WMT700 address. After WMT700 has received this command
and sent a response, the sensor does not respond to any other command for one second.
This gives WMT700 time to write the new address to the non-volatile memory.
You can also change the WMT700 address by entering
configuration mode and changing
the address parameter.
Chapter 5 – Operation
155