25-5
IM PX8000-02EN
Code Message Chapter or
Section
719 Cannot execute Time search when the time base setting is to be an External clock. 4.1
723 Cannot update null value.
No measured value, calculated by numeric, is available to set.
24.2
724 Calibration error occurred.
Restart this machine.
If it occurred again, maintenance service is required.
725 Cannot execute when GO/NO-GO traces contain an AUX channel with the sense type set to pulse. 2.4 to 2.6
726 Cannot execute when search trace contains an AUX channel with the sense type set to pulse. 2.4 to 2.6
727 Cannot execute when trigger sources contain an AUX channel with the sense type set to pulse. 2.4 to 2.6
728 Cannot execute or change while running.
Press the START/STOP key to stop acquisition.
4.2
729 No measurable channel.
Setup Errors (800 to 899)
Code Message Chapter or
Section
800 Illegal date-time.
Set the correct date and time.
801 Cannot set these file name.
- Over 32 characters.
- Include inhibit characters.
- Inhibit MS-DOS file name.
Enter an other file name.
22.3
803 Cannot change this parameter while running.
Press the START/STOP key to stop acquisition.
4.2
804 Cannot change settings during GO/NO-GO.
Stop the GO/NO-GO (Stop the Acquire).
Chapter 17
805 Can not change display points with this T/div setting. 2.8
806 Cannot be changed when trigger A is not X.
Set the state of the channel corresponding to condition A to ‘X’.
3.8 to 3.10
807 Cannot set while TimeSynchro setting not Off. 24.3
808 Cannot change when Channel Display is OFF or Math settings are invalid.
Set the channel display ON or make appropriate Math settings.
8.1, chapter
15
809 Cannot change when External Clock is active. 4.1
810 Cannot change while running. 4.2
811 Illegal math expression.
Input a correct computing equation.
15.5
812 Cannot set this model
813 Cannot set anything other than Low Pass for a Gaussian filter.
Change the Filter Type to another filter besides Gaussian.
15.5
815 Cannot change settings during Action mode.
Stop the Action.
18.1
816 Cannot set the channels which do not have modules installed. Chapter 2,
section 25.3
817 Cannot Set or Execute.
818 If the trigger mode is set to Single, Single(N), or OnStart, the acquisition mode cannot be set to Average. 3.1
819 If the acquisition mode is Average, the trigger mode cannot be set to Single, Single (N), or OnStart. 4.1
820 The acquisition mode cannot be set in the current record length. 4.1
822 Cannot be configured or executed during the search operation. Chapter 19
823 Cannot be configured or executed during the history search operation. 20.2
824 The record cannot be selected. Chapter 20
825 History record does not exist. Chapter 20
826 Cannot be configured or executed while computation is in progress.
Aborted when history display mode is set to One.
20.1
827 Cannot be configured or executed while updating the history all display.
Aborted when history display mode is set to One.
20.1
25.1 Messages and Corrective Actions