25-6
IM PX8000-02EN
Code Message Chapter or
Section
829 Zones cannot be edited in the following cases:
- When the main window is not displayed.
- When the relevant waveform is not displayed.
5.1, 8.1, 11.1,
12.1, 16.1,
17.1
830 The zone waveform does not exist. 17.1
832 Zones determination is not possible in the following cases:
- When the main window is not displayed.
- When the relevant waveform is not displayed.
- When the zone waveform does not exist.
5.1, 8.1, 11.1,
12.1, 16.1,
17.1
833 Processing statistics.
To perform other operations, abort the statistical processing.
14.2
834 The channel which couldn’t be set up was specified. —
835 Cannot be set when the acquisition mode is set to average. 4.1
836 Cannot be changed when VScale is SPAN. 8.1
838 It is an unacceptable parameter to set up to the present module. —*
840 Cannot be set to a range of 20 sec/div to 2 min/div during roll display. 2.8
842 Zooming is not available when the number of displayed points of the FFT waveform is less than 50 in the
Zoom window.
12.1, 15.5,
16.1
846 P-P compression cannot be used to save when a record length is 1K. —
852 Cannot set Math to OFF while FFT Window ON. 16.1
853 Cannot select this trace because it already selected. —
855 Cannot change to such Record length while running.
Set the trigger mode to Auto, decrease T/Div to less than 100 msec/div to enable roll mode, or set the
trigger mode to Single or On Start.
2.8, 3.1
856 Cannot Display setting to ON.
This CH didn’t acquisition to memory.
—
858 Cannot set while action mode is ON. 18.1
867 Cannot be specified when the print style is Numeric. —
869 Cannot set while GO/NO-GO mode.
Turn OFF GO/NO-GO mode first.
Chapter 17
871 No effective channel for Math Setup. —
872 No effective channel for History Search Setup. —
874 Cannot set Save Range except ‘Main’ while PP-Comp save mode. —
876 Cannot frame setting to ON, except Image format on JPEG. 21.3
877 Cannot set to display points under 100. 12.1
879 Cannot set GO/NO-GO mode while Math or FFT Window is ON. Chapter 15,
section 16.1
882 Cannot set while Single-N running. 3.1
890 Illegal math expression.
Input a correct computing equation.
7.4
891 Cannot simultaneously set current range when sense types differ.
To copy or set ranges to other current channels, first unify sense type.
2.2
892 Cannot set when current module types differ. 2.2
893 Cannot execute.
Select copy destination channels(elements).
2.7
* Communication Interface User’s Manual, IM PX8000-17EN
System Errors (900 to 999)
Code Message Chapter or
Section
900 No module installed.
Install the module.
2.3*
901 Failed to backup setup data.
Will initialize.
Backup battery may be low.
2.4*
902 The firmware is not suitable for this system.
Install the proper firmware.
—
903 The USB device’s power consumption exceeded the capacity of the USB hub. 7.9*
906 Fan stopped.
Maintenance service is required.
—
* Getting Started Guide, IM PX8000-03EN
25.1 Messages and Corrective Actions