25-7
IM PX8000-02EN
Code Message Chapter or
Section
907 Internal temperature is too high.
Maintenance service is required.
It will shutdown automatically.
—
908 Check the measured current and the number of probes that you are using. 2.12*
910 Key protect is enabled.
To release the protection, press the PROTECT key or enter the password.
24.9
911 Fan for Input modules stopped.
Cannot start.
Maintenance service is required.
—
912 Fan for CPU stopped.
Maintenance service is required.
It will shutdown automatically.
—
913 LCD BackLight Failure.
Maintenance service is required.
—
914 Cannot start while this module configuration.
760811 should use in Slot 1,Slot 3,Slot 5,Slot 7.
760812 should use in Slot 2,Slot 4,Slot 6,Slot 8.
760813 should use in Slot 2,Slot 4,Slot 6,Slot 8.
760851 should use in Slot 3,Slot 5,Slot 7.
Slot 1 and slot 2 must be installed.
2.3*
915 It installed the module which cannot support by this machine. Or it installed the module which cannot be
used in the slot.
2.3*
917 Hardware configuration error occurred.
Restart this machine.
If it occurred again, maintenance service is required.
—
918 Error occurred while ImageFile process. —
919 Key operate not available while system error occurred —
920 Firmware overwriting error occurred. —
922 Internal hardware communication error has occurred. —
923 An error has occurred in the power measurement element.
Power off immediately.
Maintenance service is required.
—
924 Unable to start because there is an error in the suffix code.
Maintenance service is required.
—
* Getting Started Guide, IM PX8000-03EN
Note
If servicing is required, first see if initializing the instrument fixes the problem.
25.1 Messages and Corrective Actions