3-26
IM 701240-01E
Immediately after connecting the thermocouple, the heat balance may be
disturbed at the input terminal section and may cause measurement errors.
Therefore, wait about 10 minutes before making a measurement.
In an environment where the air from the air conditioning is directly applied to
the input terminals or where there are effects from a heat source, the heat
balance may be disturbed at the input terminal section and cause measurement
errors.
When making measurements in this type of environment, take preventive
measures such as changing the position.
3.10 Connecting Thermocouples