Waveform Acquisition
7.2 Setting the Acquisition Mode
<For a description of this function, refer to page 2-28.>
Procedure
1. Press MENU.
2. Press the ACQ soft key.
Setting the Acquisition Mode
3. Press the Mode soft key.
4. Press the Normal, Envelope, Average, or Box Average soft key.
You may not be able to select some modes depending on the trigger mode setting.
Setting the Acquisition Count
5. Press the Count soft key.
You cannot set the acquisition count when the trigger mode is Single, Single (N), or Log.
6. Turn the jog shuttle to set the acquisition count. Pressing RESET resets the
count to Infinite.
If the acquisition mode is Average and you set Count to Infinite, proceed to step 7.
Setting the Attenuation (When the acquisition mode is Average and Count
is Infinite)
7. Press the Weight soft key.
8. Turn the jog shuttle to set the attenuation.