6-8
IM 701240-01E
6.5 Setting the Edge Trigger (SIMPLE)
<For a description of this function, refer to page 2-17.>
Procedure
1. Press MENU.
2. Press the TRIGGER soft key.
3. Press the Trigger Cond. Setup soft key.
4. Press the Select soft key to select Simple.
Setting the Trigger Source
5. Press the Source soft key. The trigger source selection menu appears.
6. Use jog shuttle+SELECT to select the channel to be used as the trigger source
from CH1 to CH16.
Note
The menu does not appear for channels without modules.
Setting the Trigger Level
7. Press the Level soft key.
8. Turn the jog shuttle to set the trigger level. Pressing RESET sets the trigger
level to 0.
Note
The trigger level setting applies to both simple and enhanced triggers.
Setting the Trigger Slope
9. Press the Slope soft key to select , , or .
Setting the Hysteresis
10. Press the Hysteresis soft key to select , , or .
Setting the Hold Off
11. Set the hold off time according to the procedures given in section 6.4.